RU2426491C2 - Method of programming sensitivity to r-wave in devices for subcutaneous monitoring of electrocardiogram - Google Patents
Method of programming sensitivity to r-wave in devices for subcutaneous monitoring of electrocardiogram Download PDFInfo
- Publication number
- RU2426491C2 RU2426491C2 RU2009127989/14A RU2009127989A RU2426491C2 RU 2426491 C2 RU2426491 C2 RU 2426491C2 RU 2009127989/14 A RU2009127989/14 A RU 2009127989/14A RU 2009127989 A RU2009127989 A RU 2009127989A RU 2426491 C2 RU2426491 C2 RU 2426491C2
- Authority
- RU
- Russia
- Prior art keywords
- muscle
- wave
- voltage
- subcutaneous
- electrocardiogram
- Prior art date
Links
Images
Abstract
Description
Изобретение относится к медицине, а именно к инвазивной аритмологии, и предназначено для длительной диагностики аритмий сердца.The invention relates to medicine, namely to invasive arrhythmology, and is intended for the long-term diagnosis of cardiac arrhythmias.
В течение последних лет усиливается интерес к длительному подкожному мониторированию электрокардиограммы (ЭКГ), что позволяет всесторонне оценить аритмический статус пациента и осуществлять динамическое наблюдение за изменением этого статуса на протяжении нескольких месяцев или лет. Аппарат для подкожного мониторирования ЭКГ представляет собой миниатюрное устройство длиной несколько сантиметров, которое под местной анестезией имплантируется в подкожную жировую клетчатку грудной клетки. После чего устройство осуществляет непрерывную запись подкожной ЭКГ и сохраняет в памяти фрагменты, которые при автоматическом анализе относит к аритмическим или которые фиксируются пациентов при помощи специального активатора.Over the past few years, interest has been growing in prolonged subcutaneous monitoring of the electrocardiogram (ECG), which allows a comprehensive assessment of the arrhythmic status of the patient and dynamic monitoring of changes in this status over several months or years. The apparatus for subcutaneous ECG monitoring is a miniature device several centimeters long, which is implanted under local anesthesia into the subcutaneous fatty tissue of the chest. After that, the device continuously records the subcutaneous ECG and saves fragments that are classified as arrhythmic during automatic analysis or which are fixed to patients using a special activator.
Важное значение имеет уровень чувствительности аппарата к R-волне (адекватность восприятия R-волн), поскольку все алгоритмы автоматического обнаружения аритмии и ее анализ базируются на частоте и степени ритмичности именно R-волн.The level of sensitivity of the apparatus to the R-wave (adequate perception of R-waves) is of great importance, since all algorithms for automatic detection of arrhythmia and its analysis are based on the frequency and degree of rhythm of R-waves.
Известен способ программирования чувствительности к R-волне в аппаратах для подкожного мониторирования ЭКГ на основе оценки подкожной ЭКГ и анализа вольтажа Р- и R-волн [1].A known method of programming sensitivity to the R-wave in the apparatus for subcutaneous ECG monitoring based on the assessment of the subcutaneous ECG and analysis of the voltage of the P and R waves [1].
Пациенту после имплантации устройства для подкожного мониторирования ЭКГ на область имплантации устанавливают головку программатора и на экран программатора выводят подкожную ЭКГ. Проводят анализ и оценку вольтажа Р-волны и R-волны, после чего осуществляют программирование уровня чувствительности (sensitivity) на значение в диапазоне от вольтажа Р-волны до вольтажа R-волны.After the implantation of the device for subcutaneous ECG monitoring, the head of the programmer is installed on the implantation area and the subcutaneous ECG is displayed on the screen of the programmer. An analysis and evaluation of the voltage of the P-wave and R-wave is carried out, after which the sensitivity level (sensitivity) is programmed to a value in the range from the voltage of the P-wave to the voltage of the R-wave.
Однако данный способ обладает существенным недостатком. А именно не учитывается степень так называемого мышечного «шума», который всегда возникает на подкожной ЭКГ во время движения рук пациента. Мышечный «шум» искажает качество подкожной ЭКГ, ложно детектируется аппаратом как серия из R-волн. А это, в свою очередь, приводит к ложной детекции аритмий.However, this method has a significant drawback. Namely, the degree of the so-called muscle “noise” that always occurs on the subcutaneous ECG during the movement of the patient’s arms is not taken into account. Muscular “noise” distorts the quality of the subcutaneous ECG and is falsely detected by the device as a series of R-waves. And this, in turn, leads to a false detection of arrhythmias.
Целью данного изобретения является улучшение качества диагностики аритмий в аппаратах для подкожного мониторирования электрокардиограммы. Поставленная цель достигается тем, что уровень чувствительности к R-волне программируется в соответствии с уровнем возможного мышечного «шума».The aim of this invention is to improve the quality of diagnosis of arrhythmias in devices for subcutaneous monitoring of an electrocardiogram. The goal is achieved in that the level of sensitivity to the R-wave is programmed in accordance with the level of possible muscle “noise”.
Способ осуществляют следующим образом.The method is as follows.
Пациенту после имплантации устройства для подкожного мониторирования ЭКГ на область имплантации устанавливают головку программатора 1 и на экран программатора 2 выводят подкожную ЭКГ 3, что представлено на фиг.1.After the implantation of the device for subcutaneous ECG monitoring, the head of the programmer 1 is installed on the implantation area and the
Осуществление мышечной пробы 4 (отображено на фиг.1) выполняют следующим образом: добиваются статического напряжения большой грудной мышцы за счет того, что пациент пытается провести движение в плечевом суставе на стороне имплантации устройства, а врач в это время оказывает противодействие данному движению. Таким образом, создается имитация мощной физической нагрузки, которую пациент может выполнять в бытовых условиях и которая может приводить к формированию мышечного «шума».The implementation of muscle samples 4 (shown in figure 1) is performed as follows: they achieve static tension of the pectoralis major muscle due to the fact that the patient is trying to move in the shoulder joint on the side of the implantation of the device, and the doctor at the same time counteracts this movement. Thus, an imitation of a powerful physical load is created, which the patient can perform in the domestic environment and which can lead to the formation of muscle “noise”.
Записывают подкожную ЭКГ 3 во время мышечной пробы, когда производится попытка сгибания, разгибания, приведения и отведения в плечевом суставе.The
Затем проводят измерение вольтажа R-волны и вольтажа мышечного «шума», после чего осуществляют программирование уровня чувствительности (sensitivity) на значение в диапазоне от наибольшего вольтажа мышечного «шума» до наименьшего вольтажа R-волны.Then, the voltage of the R-wave and the voltage of the muscle “noise” are measured, after which the sensitivity level (sensitivity) is programmed to a value in the range from the highest voltage of the muscle “noise” to the lowest voltage of the R-wave.
Применение способа иллюстрируется следующим примером.The application of the method is illustrated by the following example.
Пример. Больная К., 62 лет. В возрасте 55 лет появились приступы учащенного неритмичного сердцебиения. На электрокардиограмме в это время регистрировалась фибрилляция предсердий. Вопреки постоянному приему антиаритмических препаратов (соталол, кордарон, пропанорм) в течение последних лет приступы аритмии участились. Последний эпизод продолжался на протяжении 7 мес.Example. Patient K., 62 years old. At the age of 55 years, attacks of rapid irregular heartbeat appeared. Atrial fibrillation was recorded on the electrocardiogram at this time. Despite the constant use of antiarrhythmic drugs (sotalol, cordarone, propanorm), arrhythmia attacks have become more frequent in recent years. The last episode lasted for 7 months.
Клинический диагноз: Артериальная гипертензия III ст. Риск - 4. Хроническая фибрилляция предсердий, тахиформа (7 мес). ХСН-2А. ФК - II NYHA.Clinical diagnosis: Arterial hypertension III tbsp. Risk - 4. Chronic atrial fibrillation, tachyform (7 months). CHF-2A. FC - II NYHA.
Учитывая симптомность аритмии и ее рефрактерность к приему антиаритмических препаратов, в августе 2008 года пациентке выполнена катетерная радиочастотная аблация ганглионарных сплетений левого предсердия. Вторым этапом пациентке было имплантировано устройство для подкожного мониторирования электрокардиограммы «Reveal XT» с целью объективной оценки эффекта оперативного лечения.Given the symptom of arrhythmia and its refractoriness to taking antiarrhythmic drugs, in August 2008 the patient underwent catheter radiofrequency ablation of the ganglion plexuses of the left atrium. The second stage, the patient was implanted with a device for subcutaneous monitoring of the electrocardiogram "Reveal XT" in order to objectively assess the effect of surgical treatment.
После имплантации устройства проводилось программирование чувствительности аппарата к R-волне. Вольтаж R-волны составил 0,4 мВ, вольтаж Р-волны составил 0,06 мВ. В связи с этим уровень чувствительности был запрограммирован на значение чуть выше вольтажа Р-волны, а именно 0,05 мВ. При повторном осмотре пациентки в октябре 2008 года и опросе устройства - зафиксировано 876 эпизодов, которые были интерпретированы аппаратом как фибрилляция предсердий. Но при просмотре всех сохраненных ЭКГ выяснилось, что 823 эпизода (94%) обусловлены ложноположительным анализом в результате неверной интерпретацией мышечного «шума», т.к. высоковольтажный «шум» детектировался как множество R-волн.After implantation of the device, the sensitivity of the apparatus to the R-wave was programmed. The voltage of the R-wave was 0.4 mV, the voltage of the R-wave was 0.06 mV. In this regard, the sensitivity level was programmed to a value slightly higher than the voltage of the P-wave, namely 0.05 mV. When the patient was examined again in October 2008 and the device was interrogated, 876 episodes were recorded, which were interpreted by the device as atrial fibrillation. But when viewing all stored ECGs, it turned out that 823 episodes (94%) were caused by a false positive analysis as a result of an incorrect interpretation of muscle “noise”, because high voltage “noise” was detected as a set of R-waves.
Пациентке были проведены мышечные пробы по предложенному нами способу. На фиг.2 показаны направления мышечной пробы: разгибание (фиг.2, 5), отведение (фиг.2, 6), сгибание (фиг.2, 7), приведение (фиг.2, 8). Соответствующими цифрами обозначена подкожная электрограмма с мышечным «шумом» во время данной пробы.The patient underwent muscle tests according to our proposed method. Figure 2 shows the directions of the muscle sample: extension (figure 2, 5), abduction (figure 2, 6), flexion (figure 2, 7), reduction (figure 2, 8). The corresponding numbers indicate the subcutaneous electrogram with muscle “noise” during this test.
Результаты измерения вольтажа R-зубца и вольтажа мышечного «шума» представлены в таблице 1.The results of measuring the voltage of the R-wave and the voltage of muscle “noise” are presented in table 1.
Таблица показывает, что наименьшая амплитуда R-волны составила 0,4 мВ, а наибольший вольтаж «шума» - 0,16 мВ. Таким образом, значение чувствительности к R-волне необходимо выбирать в диапазоне между этими цифрами. Пациентке было запрограммировано значение 0,2 мВ, то есть порог чувствительности был увеличен в 4 раза (с 0,05 мВ).The table shows that the smallest R-wave amplitude was 0.4 mV, and the highest “noise” voltage was 0.16 mV. Thus, the value of sensitivity to the R-wave must be selected in the range between these figures. The patient was programmed with a value of 0.2 mV, that is, the sensitivity threshold was increased 4 times (from 0.05 mV).
При повторном осмотре пациентки в феврале 2009 года и опросе устройства - зафиксировано лишь шесть новых эпизодов фибрилляции предсердий. При просмотре всех сохраненных ЭКГ выяснилось, что все эпизоды детектированы верно, ложноположительная детекция больше не наблюдалась. Использование предложенного нами способа позволило устранить ложную интерпретацию мышечного «шума» в виде аритмических событий.When re-examining the patient in February 2009 and questioning the device, only six new episodes of atrial fibrillation were recorded. When viewing all stored ECGs, it turned out that all episodes were detected correctly, false-positive detection was no longer observed. Using our proposed method allowed us to eliminate the false interpretation of muscle “noise” in the form of arrhythmic events.
Нами проведен сравнительный анализ двух способов программирования чувствительности к R-волне в устройствах подкожного мониторирования ЭКГ:We carried out a comparative analysis of two methods of programming sensitivity to the R-wave in devices for subcutaneous ECG monitoring:
Способ №1: соответствует описанному аналогу.Method No. 1: corresponds to the described analogue.
Способ №2: предлагаемый нами.Method number 2: offered by us.
Таблица 2 убедительно показывает, что использование предложенного нами способа позволяет сократить ложноположительную детекцию в 6,5 раз и за счет этого увеличить периодичность осмотра пациента для перепрограммирования устройства в 1,9 раза.Table 2 convincingly shows that the use of the proposed method allows us to reduce false-positive detection by 6.5 times and thereby increase the frequency of examination of the patient to reprogram the device by 1.9 times.
Таким образом, использование описанного нами способа позволило повысить точность, качество и информативность автоматической диагностики аритмий в аппаратах для подкожного мониторирования ЭКГ.Thus, the use of the method described by us made it possible to increase the accuracy, quality, and information content of the automatic diagnosis of arrhythmias in devices for subcutaneous ECG monitoring.
Способ прост в применении и может быть использован в любом лечебном учреждении, имеющем в своем штате специалистов, занимающихся проблемой аритмий сердца.The method is easy to use and can be used in any medical institution that has in its staff specialists involved in the problem of cardiac arrhythmias.
ЛитератураLiterature
1. REVEAL® XT 9529: Insertable Cardiac Monitor, Clinician Manual. - 2007. - 94 p.1. REVEAL® XT 9529: Insertable Cardiac Monitor, Clinician Manual. - 2007. - 94 p.
Claims (1)
Priority Applications (1)
Application Number | Priority Date | Filing Date | Title |
---|---|---|---|
RU2009127989/14A RU2426491C2 (en) | 2009-07-20 | 2009-07-20 | Method of programming sensitivity to r-wave in devices for subcutaneous monitoring of electrocardiogram |
Applications Claiming Priority (1)
Application Number | Priority Date | Filing Date | Title |
---|---|---|---|
RU2009127989/14A RU2426491C2 (en) | 2009-07-20 | 2009-07-20 | Method of programming sensitivity to r-wave in devices for subcutaneous monitoring of electrocardiogram |
Publications (2)
Publication Number | Publication Date |
---|---|
RU2009127989A RU2009127989A (en) | 2011-01-27 |
RU2426491C2 true RU2426491C2 (en) | 2011-08-20 |
Family
ID=44755922
Family Applications (1)
Application Number | Title | Priority Date | Filing Date |
---|---|---|---|
RU2009127989/14A RU2426491C2 (en) | 2009-07-20 | 2009-07-20 | Method of programming sensitivity to r-wave in devices for subcutaneous monitoring of electrocardiogram |
Country Status (1)
Country | Link |
---|---|
RU (1) | RU2426491C2 (en) |
Cited By (1)
Publication number | Priority date | Publication date | Assignee | Title |
---|---|---|---|---|
RU2655443C2 (en) * | 2012-12-14 | 2018-05-28 | Конинклейке Филипс Н.В. | System and method to detect significant arrhythmic events through a photoplethysmogram (ppg) and accelerometer |
-
2009
- 2009-07-20 RU RU2009127989/14A patent/RU2426491C2/en not_active IP Right Cessation
Non-Patent Citations (2)
Title |
---|
REVEAL® XT 9529: Insertable Cardiac Monitor. Clinician Manual., 2007, 94 p. * |
Зотов Д.Д., Гротова А.В. Современные методы функциональной диагностики в кардиологии. - СПб.: Фолиант, 2002, с.9-50. * |
Cited By (1)
Publication number | Priority date | Publication date | Assignee | Title |
---|---|---|---|---|
RU2655443C2 (en) * | 2012-12-14 | 2018-05-28 | Конинклейке Филипс Н.В. | System and method to detect significant arrhythmic events through a photoplethysmogram (ppg) and accelerometer |
Also Published As
Publication number | Publication date |
---|---|
RU2009127989A (en) | 2011-01-27 |
Similar Documents
Publication | Publication Date | Title |
---|---|---|
Crawford et al. | ACC/AHA guidelines for ambulatory electrocardiography: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee to Revise the Guidelines for Ambulatory Electrocardiography) developed in collaboration with the North American Society for Pacing and Electrophysiology | |
JP5801866B2 (en) | Method and apparatus for assessing cardiac function | |
Romero et al. | Depolarization changes during acute myocardial ischemia by evaluation of QRS slopes: standard lead and vectorial approach | |
Karunadas et al. | Comparison of arrhythmia detection by conventional Holter and a novel ambulatory ECG system using patch and Android App, over 24 h period | |
Lauschke et al. | New implantable cardiac monitor with three-lead ECG and active noise detection | |
US20200170526A1 (en) | Detection and monitoring using high frequency electrogram analysis | |
Rishniw et al. | Effect of body position on the 6‐lead ECG of dogs | |
Haeberlin et al. | The optimal lead insertion depth for esophageal ECG recordings with respect to atrial signal quality | |
US9254094B2 (en) | Detection and monitoring using high frequency electrogram analysis | |
US20050010122A1 (en) | Spatial heterogeneity of repolarization waveform amplitude to assess risk of sudden cardiac death | |
RU2426491C2 (en) | Method of programming sensitivity to r-wave in devices for subcutaneous monitoring of electrocardiogram | |
US10292627B2 (en) | Method and apparatus for noninvasive detection of potential symptomless blood loss | |
Chrysostomakis et al. | Implantable loop recorder undersensing mimicking complete heart block | |
Drew | Standardization of electrode placement for continuous patient monitoring: introduction of an assessment tool to compare proposed electrocardiogram lead configurations | |
Flanders et al. | The next heartbeat: Creating dynamic and histographic Poincare plots for the assessment of cardiac rhythms | |
Trägårdh-Johansson et al. | Similarity of ST and T waveforms of 12-lead electrocardiogram acquired from different monitoring electrode positions | |
Zenger et al. | Pharmacological and simulated exercise cardiac stress tests produce different ischemic signatures in high-resolution experimental mapping studies | |
Sunkak et al. | Does transcatheter ventricular septal defect closure affect heart rate variability in children? | |
US20240138744A1 (en) | Apparatus and method for analysis and monitoring of high frequency electrograms and electrocardiograms in various physiological conditions | |
Nelwan et al. | Derived 12-lead ECG systems | |
Kot et al. | Analysis of the Biological Signal for Automated Diagnostics | |
Matsui et al. | A case report of undetected cardiac arrest in a patient with an insertable cardiac monitor | |
US7957787B2 (en) | Method of examining dynamic cardiac electromagnetic activity and detection of cardiac functions using results thereof | |
Dori et al. | Non-invasive computerised detection of acute coronary occlusion | |
Kossick et al. | PhD" Linda Hill, DNP"(" DNS, Union University, Jackson, TN, USA:" University of Alabama at Birmingham, Birmingham, AL, USA;“University |
Legal Events
Date | Code | Title | Description |
---|---|---|---|
MM4A | The patent is invalid due to non-payment of fees |
Effective date: 20120721 |