The content of the invention
In order to verify whether 3 force-measuring type centroid measurement platform precision reach its nominal precision, the present inventor is whole with multistage
Calibration technique demarcates barycenter test desk, and with 8 Calibration Technologies, torque accuracy determination method, weighing precision judgement side
Method examines equipment precision, so as to complete the present invention.
Present invention employs following technical solutions:
3 force-measuring type centroid measurement platform multistage integral calibrating methods of the present invention, comprise the following steps:
1) after in 3 force-measuring type centroid measurement platforms, installation is complete, first order calibration uses the standard test weight through metering, profit
Positioned with central shaft, be placed on centroid measurement platform, calculate respectively the respective code value increment of three sensors and standard test weight and in
The gross mass of mandrel (is denoted as m1) Mass Distribution on a sensor ratio, be less than m as when test specimen quality1When each sensor
Calibration coefficient;It is equally positioned using central shaft using the standard test weight by metering when the second level is demarcated, is placed on the first order
On standard test weight, the gross mass of the respective code value increment of three sensors and second level standard test weight and central shaft is calculated respectively
(it is denoted as m2) Mass Distribution on a sensor ratio, be more than m as when test specimen quality1Less than m1+m2When each sensor mark
Determine coefficient;And so on, calculate the 6th grade of calibration coefficient of sensor;
2) on the basis of the quality and eccentric moment that measure all standard test weights, with a standard rider, (quality is denoted as m0),
It is placed sequentially on standard test weight eight and is distributed on radius as r0In calibration hole on reference circle, eight position riders are measured respectively
The eccentric moment of generation;
3) with the eccentric moment measured value M of rideri(i=1,2,3 ..., 8) in 0.88m0r0To 1.12m0r0Between, sentence
Fixed 8 points of calibrating eccentric moments size measurement is qualified, i.e., using 8 points of calibrating hole centers as the center of circle, radius r=0.12*r0For radius
A circumference is drawn, does two tangent lines to the circumference from standard rotor center of circle O for starting point, the angle between two tangent lines is torque side
To deviation allowed band, that is, torque direction is measured in αiIt is qualified in the range of ± 6.8 °;If torque size and direction are qualified,
And m0r0Less than centroid measurement platform nominal torque accuracy, then its torque accuracy qualification is judged;
4) to choose the standard test weight of 10%~20% mass of centroid measurement platform range, duplicate measurements three times, measured value with
Standard test weight metered mass maximum measured deviation is denoted as Δ mi(i=1,2,3);To choose centroid measurement platform range 80%~90%
The standard test weight of quality, three times, measured value is denoted as Δ m with standard test weight metered mass maximum measured deviation for duplicate measurementsi(i=
4,5,6), if Δ mi(i=1 ... 6) is respectively less than the weighing precision of centroid measurement platform nominal, then judges its weighing precision qualification.
Wherein, αi=0 ° of+n × 45 ° n=0,1,2,3,4,5,6,7.
The present invention had not only considered demarcates single sensor ignored coupling factor influence respectively, but also avoids confronting
Heart test desk frequently dismounts, so as to improve calibration calibration accuracy and efficiency.
Specific embodiment
The specific embodiment of the present invention is described in detail below, which is exemplary, not
It is intended to carry out any restriction to protection scope of the present invention.
3 force-measuring type centroid measurement platform multistage integral calibrating methods of the present invention are for existing 3 force-measuring type matter
What heart test desk carried out, including multiple demarcating steps, specifically include procedure below:
Multistage entirety coefficient calibration:
1. measuring barycenter platform zero-bit, three sensor code values are respectively 0,0,0.
2. demarcate first order coefficient.
A) first order standard rotor is lifted into upper barycenter platform, central shaft positioning.
B) computer input rotor standard quality (axis containing centralized positioning) standard value m1=3000kg.
C) computer reads each sensor code value variation after instrument indicating value to be tested is stablized, and is respectively 992,990,1002.
D) each transducer calibration coefficient of the first order is calculated, is respectively 1.008064516 (=m13 ÷ 992 of ÷),
1.01010101 (=m13 ÷ 990 of ÷), 0.998003992 (=m1÷3÷1002)。
3. demarcate second level coefficient.
A) second level standard rotor is lifted on first order rotor, central shaft positioning.
B) rotor quality (axis containing centralized positioning) standard value m is inputted2=3003kg.
C) computer reads each sensor code value variation after instrument indicating value to be tested is stablized, and is respectively 997,993,1005.
D) each transducer calibration coefficient in the second level is calculated, is respectively 1.004012036 (=m23 ÷ 997 of ÷),
1.008056395 (=m23 ÷ 993 of ÷), 0.9960199 (=m2÷3÷1005)。
4. it is carried out successively until the 6th grade.
5. nominal data is deposited and printed.
6. all standard rotors are hung into barycenter platform.
8 Calibration Technologies:
A. computer enters 8 validation programs, measures barycenter platform zero-bit.Three sensor code values are respectively 0,0,0.
B. two standard rotor m are chosen1、m2, gross weight 6003kg, weight weighs the 10% of full scale in equipment
(1700kg) in the range of 90% (15300kg), barycenter platform central shaft positions in lifting.
C. rotor quadrant is aligned with barycenter platform quadrant.
D. it is m to choose a quality0The rider of=0.4kg, travelling radius r0=0.5m by standard rotor upper Positioning holes position
Put decision so that m0r0≤ M (centroid measurement platform nominal torque accuracy, 0.25kgm).
E. the quality of measurement standard rotor and eccentric moment value, deposit after being measured.
F. successively by rider be inserted into standard rotor on reference radius be 0.5m, positioned at 0 °, 45 °, 90 °, 135 °, 180 °,
The location hole of eight on 225 °, 270 °, 315 °, and eccentric moment M is measured respectivelyi(i=1,2,3 ..., 8) are respectively M1=
0.194kgm, M2=0.198kgm, M3=0.212kgm, M4=0.203kgm, M1=0.198kgm, M1=0.189kgm, M1=
0.182kgm, M1=0.187kgm.Torque accuracy determination method:
A. eccentric moment measured value Mi(i=1,2,3 ..., 8) are in 0.88m0r0(=0.176) is to 1.12m0r0(=
0.224) between, 8 points of calibrating eccentric moment size measurements are qualified.
B. eccentric moment direction is respectively 4.5 °, 44.8 °, 90.2 °, 134.0 °, 179.3 °, 225.6 °, 269.8 °,
315.9 °, positioned in αi±6.8°(αi=0 °, 45 °, 90 °, 135 °, 180 °, 225 °, 270 °, 315 °) in the range of, 8 points of calibratings
Eccentric moment orientation measurement is qualified.
C. all qualified then 8 points of calibratings are qualification to torque size and Orientation.
Weighing precision determination method:
A. one or a set of standard rotor is chosen, weight 3000kg weighs the 10%~20% of full scale in equipment, into
Row is weighed, and is repeated 3 times, and records measured value, is respectively 3000.1kg, 2999.8kg, 3000.2kg.Calculate measured value and standard value
Error amount Δ mi(i=1,2,3).
B. one or a set of standard rotor is chosen, weight 15003kg weighs full scale 80%~90% in equipment, into
Row is weighed, and is repeated 3 times, and records measured value, is respectively 15002.8kg, 15003.1kg, 15003.3kg.Calculate measured value and mark
The error amount Δ m of quasi- valuei(i=4,5,6).
C. as Δ mi(i=1 ... 6)≤Δ m (weighing precision of centroid measurement platform nominal is herein 0.5kg), weighing precision
It is qualified.
Technical scheme is described in detail in embodiments described above, it should be understood that described above
Only specific embodiments of the present invention are not intended to limit the invention, it is all done in the spirit of the present invention any repair
Change, supplement or similar fashion substitute etc., should be included within the scope of the present invention.