Summary of the invention
Technical matters to be solved by this invention is, overcomes the defective of above-mentioned prior art and the assay method of objectionable impurities caprolactam content in a kind of workplace air is provided.
The assay method of objectionable impurities caprolactam in the workplace air of the present invention is to utilize liquid chromatography-tandem mass spectrometry combined instrument (LC-MS/MS) to detect.
The assay method of objectionable impurities caprolactam in the above-mentioned workplace air, its concrete steps are as follows:
(1) sample collecting, transportation and preservation part
Two impinger of adorning 10.0 mL water in each of connecting, sealing absorption tube port is put into sales kit (SK) with absorption tube and is brought to sampled point; Take out absorption tube and be connected,, write down temperature, air pressure and the humidity of sampled point simultaneously with 15 minutes air samples of 3 L/min traffic samplings with atmosphere sampling instrument;
After the sampling, seal the air inlet/outlet of absorption tube immediately, uprightly place transportation and preservation in the cleaning container.Should analyze as early as possible behind the sample collecting,, should under 4 ℃ of conditions, keep in Dark Place, be no more than 7 days as can not in time analyzing;
(2) analysis part of sample
The absorption of adopting sample is washed the draft tube inwall 3 times of absorption tube night, after shaking up, to absorb and pour into night in the tool plug test tube, adopt disposable sample introduction needle to suck about 1 mL sample, organic phase pin type filter through 0.2 μ m filters directly injection liquid phase sample introduction bottle of back, detects with the liquid chromatography-tandem mass spectrometry combined instrument.If the concentration of determinand surpasses measurement range in the sample liquid, available absorption dilution at night back is measured, and multiply by extension rate during calculating;
In the step (2), the Ultra Performance Liquid Chromatography instrument be Acquity UPLC (waters company, USA), tandem mass spectrometer be the triple level Four bar of Quattro Premier XE mass spectrum (waters company, USA);
Liquid phase chromatogram condition is:
ACQUITY UPLC BEHC18 reverse-phase chromatographic column (2.1 mm * 50 mm * 1.7 μ m);
Column temperature: 40-60 ℃;
Sample size: 3-10 μ L;
Flow rate of mobile phase: 0.3-0.6 mL/min;
Moving phase is made up of organic phase and inorganic phase solvent, and organic phase can be methyl alcohol, formic acid, acetone, second eyeball etc., also can be that wherein two or more mixed solution is formed, and inorganic is water mutually;
The moving phase type of elution can adopt isocratic elution or linear gradient elution;
The tandem mass spectrum condition:
Adopt the ionization of electron spray positively ionized pattern;
Multipole ionic reaction detection mode;
Source temperature: 100 ℃-300 ℃;
Taper hole voltage: 20-80 V;
Collision energy: 10-50 V;
Mass spectrum is optimized: Q1 full scan scope is 90-130, and unit resolution rate sweep time is 0.2 s.The collision energy that adopts in the MRM pattern is 17 ev, and the quantitative reaction passage is m/z 113.5-69, and the qualitative reaction passage is 113.5-79, and single reaction channel sweep time is 0.2 s;
Concrete mass spectrum condition is as follows:
Parameter |
Optimised value |
Capillary voltage? (V) |
3.0 |
Cone voltage? (V) |
38 |
Source temperature? (℃ ) |
100 |
Desolvation temperature? (℃) |
400 |
Desolvation gas flow? (L/h ) |
700 |
Cone gas flow? (L/h ) |
50 |
Collision energy? (ev) |
17 |
Collision Gas flow(ml/min) |
0.8 |
(3) drafting of typical curve
Accurately take by weighing a certain amount of caprolactam solid standard specimen, be dissolved in the water, quantitatively be transferred in the volumetric flask, and be diluted to scale, this solution is 1000 μ g/mL caprolactam standard stock solutions.Face with before, become the standard series of 5 concentration gradients with 1:9 methanol solvent dilution, the liquid chromatography-tandem mass spectrometry combined instrument is adjusted to the optimum determining state, measure each standard series; Each concentration replication 3 times, with peak area to caprolactam concentration (μ g/L) drawing standard curve; Coefficient R〉0.9990;
(4) calculating section
With the operating conditions working sample and the blank solution of bioassay standard series, the sample peak area of measuring deducted the peak area of blank after, get the concentration of caprolactam by typical curve, computation process is:
By formula (1) sampling volume is converted into the standard sample volume
In the formula:
Vo-standard sample volume, L;
V-sampling volume, L;
The temperature of t-sampled point, ℃;
The atmospheric pressure of P-sampled point, kPa;
Calculate the concentration of caprolactam in the air by formula (2):
In the formula:
The concentration of caprolactam in C-air, mg/ m
3
The volume of v-sample solution, ml;
The concentration of the caprolactam of c-record, μ g/ml;
Vo-standard sample volume, L.
With prior art relatively, advantage of the present invention is:
(1) adopt liquid chromatography-tandem mass spectrometry combined instrument of the present invention to carry out the detection of objectionable impurities caprolactam in the workplace air, compare existing gas chromatographic technique and high-efficient liquid phase chromatogram technology, this analytical approach fast, accurate and effective, interference be little, precision and accuracy are all less than 10%, detect and be limited to 0.08 μ g/L, concentration limit is 1.8 * 10
-5Mg/m
3(to gather the 45L air sample);
(2) by selecting suitable moving phase, control suitable flow velocity, can make caprolactam in the sample carry out optimization from;
(3) by selecting multipole ionic reaction detection mode, method selectivity height, qualitative accurately, detection limit is low.
This shows that the method among the present invention is the mensuration of objectionable impurities caprolactam content in the workplace air, provide a kind of not only reliable but also accurate feasible method, the needs that can satisfy research and produce.
Embodiment
Embodiment 1, the mensuration of caprolactam in certain plastic molding and processing plant's polymerization plant air.
The polymerization plant sampled point (temperature: 20 ℃, air pressure: 100.3 kPa), take out absorption tube and be connected, with 15 minutes air samples of 3 L/min traffic samplings with atmosphere sampling instrument.After the sampling, seal the air inlet/outlet of absorption tube immediately, uprightly place in the cleaning container.The absorption of adopting sample is washed the draft tube inwall 3 times of absorption tube night, after shaking up, to absorb and pour into night in the tool plug test tube, adopt disposable sample introduction needle to suck about 1 mL sample, organic phase pin type filter through 0.2 μ m filters directly injection liquid phase sample introduction bottle of back, detects with the liquid chromatography-tandem mass spectrometry combined instrument.Do blank determination with 1:9 methanol solvent simultaneously.If the concentration of determinand surpasses measurement range in the sample liquid, available absorption dilution at night back is measured, and multiply by extension rate during calculating.
(3) drafting of typical curve
Accurately take by weighing 0.1000g caprolactam solid standard specimen, be dissolved in the water, quantitatively be transferred in the 100 mL volumetric flasks, and be diluted to scale, this solution is 1000 μ g/mL standard stock solutions.Before facing usefulness, become the standard working curve of 5 concentration gradients (10,20,40,80,160 μ g/L) stand-by with 1:9 methanol solvent dilution.The liquid chromatography-tandem mass spectrometry combined instrument is adjusted to the optimum determining state, measures each standard series; Each concentration replication 3 times, with peak area to caprolactam concentration (μ g/L) drawing standard curve; Coefficient R〉0.9990.
Detecting instrument is the liquid chromatography-tandem mass spectrometry combined instrument, wherein the Ultra Performance Liquid Chromatography instrument be Acquity UPLC (waters company, USA), tandem mass spectrometer is the triple level Four bar of a Quattro Premier XE mass spectrum (waters company, USA), testing conditions is:
Chromatographic column: ACQUITY UPLC BEH C18 (2.1 mm * 50 mm * 1.7 μ m);
Column temperature: 40 ℃;
Sample size: 5 μ L;
Flow rate of mobile phase: 0.3 mL/min;
Moving phase is formed: gradient elution, and concrete eluent gradient is as shown in table 1, and wherein, A is methyl alcohol mutually, and B is 0.1% formic acid/aqueous solution mutually;
Time(min) |
A |
B | CURVE | |
0 |
10 |
90 |
INITIAL |
3 |
30 |
70 |
6 |
5 |
100 |
0 |
1 |
8 |
10 |
90 |
1 |
The tandem mass spectrum condition:
Ion gun: electron spray ESI, positive ion;
Scan mode: multiple-reaction monitoring MRM;
Source temperature: 100 ℃-300 ℃;
Taper hole voltage: 20-80 V;
Collision energy: 10-50 V;
Mass spectrum is optimized: Q1 full scan scope is 90-130, and unit resolution rate sweep time is 0.2 s.The collision energy that adopts in the MRM pattern is 17 ev, and the quantitative reaction passage is m/z 113.5-69, and the qualitative reaction passage is 113.5-79, and single reaction channel sweep time is 0.2 s;
Concrete mass spectrum condition:
Parameter |
Optimized value |
Capillary voltage? (V) |
3.0 |
Cone voltage? (V) |
38 |
Source temperature? (℃ ) |
100 |
Desolvation temperature? (℃) |
400 |
Desolvation gas flow? (L/h ) |
700 |
Cone gas flow? (L/h ) |
50 |
Collision energy? (ev) |
17 |
Collision Gas flow (ml/min) |
0.8 |
Qualitative according to retention time, the peak area external standard method is quantitative.
(4) calculating section
With the operating conditions working sample and the blank solution of bioassay standard series, the sample peak area of measuring deducted the peak area of blank after, get the concentration of caprolactam by typical curve, the computing formula of caprolactam content is in the sample:
By formula (1) sampling volume is converted into the standard sample volume
In the formula:
Vo-standard sample volume, L;
V-sampling volume, L;
The temperature of t-sampled point, ℃;
The atmospheric pressure of P-sampled point, kPa;
Calculate the concentration of caprolactam in the air by formula (2):
In the formula:
The concentration of caprolactam in C-air, mg/ m
3
The volume of v-sample solution, ml;
The concentration of the caprolactam of c-record, μ g/ml;
Vo-standard sample volume, L.
After this plastic sheeting workshop air sample repeated test 6 times, calculate by above-mentioned formula, by testing result relative standard deviation<10% as can be known.
? |
1 |
2 |
3 |
4 |
5 |
6 |
Mean value |
RSD |
The result (10
-4 mg/m
3)
|
9.48 |
9.45 |
9.46 |
9.46 |
9.45 |
9.46 |
9.46 |
0.12% |
Embodiment 2, the mensuration of caprolactam in certain chemical fibre factory's blowing workshop air.
Blowing workshop sampled point (temperature: 22 ℃, air pressure: 99.3 kPa), take out absorption tube and be connected, with 15 minutes air samples of 3 L/min traffic samplings with atmosphere sampling instrument.After the sampling, seal the air inlet/outlet of absorption tube immediately, uprightly place in the cleaning container.The absorption of adopting sample is washed the draft tube inwall 3 times of absorption tube night, after shaking up, to absorb and pour into night in the tool plug test tube, adopt disposable sample introduction needle to suck about 1 mL sample, organic phase pin type filter through 0.2 μ m filters directly injection liquid phase sample introduction bottle of back, detects with the liquid chromatography-tandem mass spectrometry combined instrument.Do blank determination with 1:9 methanol solvent simultaneously.If the concentration of determinand surpasses measurement range in the sample liquid, available absorption dilution at night back is measured, and multiply by extension rate during calculating.
(3) drafting of typical curve
Accurately take by weighing 0.1000 g caprolactam solid standard specimen, be dissolved in the water, quantitatively be transferred in the 100 mL volumetric flasks, and be diluted to scale, this solution is 1000 μ g/mL standard stock solutions.Before facing usefulness, become the standard working curve of 5 concentration gradients (1,2,5,10,20 μ g/L) stand-by with 1:9 methanol solvent dilution.The liquid chromatography-tandem mass spectrometry combined instrument is adjusted to the optimum determining state, measures each standard series; Each concentration replication 3 times, with peak area to caprolactam concentration (μ g/L) drawing standard curve, coefficient R 0.9990.
Detecting instrument is the liquid chromatography-tandem mass spectrometry combined instrument, wherein the Ultra Performance Liquid Chromatography instrument be Acquity UPLC (waters company, USA), tandem mass spectrometer is the triple level Four bar of a Quattro Premier XE mass spectrum (waters company, USA), testing conditions is:
Chromatographic column: ACQUITY UPLC BEH C18 (2.1 mm * 50 mm * 1.7 μ m);
Column temperature: 40 ℃;
Sample size: 5 μ L;
Flow rate of mobile phase: 0.3 mL/min;
Moving phase is formed: gradient elution, and concrete eluent gradient is as shown in table 1, and wherein, A is methyl alcohol mutually, and B is 0.1% formic acid/aqueous solution mutually;
Time(min) |
A |
B | CURVE | |
0 |
10 |
90 |
INITIAL |
3 |
30 |
70 |
6 |
5 |
100 |
0 |
1 |
8 |
10 |
90 |
1 |
The tandem mass spectrum condition:
Ion gun: electron spray ESI, positive ion;
Scan mode: multiple-reaction monitoring MRM;
Source temperature: 100 ℃-300 ℃;
Taper hole voltage: 20-80 V;
Collision energy: 10-50 V;
Mass spectrum is optimized: Q1 full scan scope is 90-130, and unit resolution rate sweep time is 0.2 s.The collision energy that adopts in the MRM pattern is 17 ev, and the quantitative reaction passage is m/z 113.5-69, and the qualitative reaction passage is 113.5-79, and single reaction channel sweep time is 0.2 s;
Concrete mass spectrum condition:
Parameter |
Optimized value |
Capillary voltage? (V) |
3.0 |
Cone voltage? (V) |
38 |
Source temperature? (℃ ) |
100 |
Desolvation temperature? (℃) |
400 |
Desolvation gas flow? (L/h ) |
700 |
Cone gas flow? (L/h ) |
50 |
Collision energy? (ev) |
17 |
Collision Gas flow (ml/min) |
0.8 |
Qualitative according to retention time, the external standard peak area method is quantitative.
(4) calculating section
With the operating conditions working sample and the blank solution of bioassay standard series, the sample peak area of measuring deducted the peak area of blank after, get the concentration of caprolactam by typical curve, the computing formula of caprolactam content is in the sample:
By formula (1) sampling volume is converted into the standard sample volume
In the formula:
Vo-standard sample volume, L;
V-sampling volume, L;
The temperature of t-sampled point, ℃;
The atmospheric pressure of P-sampled point, kPa;
Calculate the concentration of caprolactam in the air by formula (2):
In the formula:
The concentration of caprolactam in C-air, mg/ m
3
The volume of v-sample solution, ml;
The concentration of the caprolactam of c-record, μ g/ml;
Vo-standard sample volume, L.
After this chemical fibre factory's blowing workshop air sample repeated test 6 times, calculate by above-mentioned formula, by testing result relative standard deviation<10% as can be known.
? |
1 |
2 |
3 |
4 |
5 |
6 |
Mean value |
RSD |
The result (10
-4 mg/m
3)
|
1.42 |
1.43 |
1.41 |
1.40 |
1.44 |
1.41 |
1.42 |
1.04% |
Obviously, those skilled in the art, can constitute assay method with of the present invention a kind of by the mensuration of liquid chromatography-tandem mass spectrometry combined instrument to objectionable impurities caprolactam content in the workplace air to objectionable impurities caprolactam content in all kinds workplace air.
The foregoing description is only for the usefulness that the present invention is described; and be not to be limitation of the present invention; the those of ordinary skill in relevant technologies field; without departing from the present invention; can also make various variations and modification; therefore all technical schemes that are equal to also should belong to category of the present invention, and scope of patent protection of the present invention should be limited by each claim.